Status alarms and Status Alarm Severity (continued)Table 5-2:
Alarm code Status message Default severity Notes Configurable?
A114 mA Output 2 Fixed
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A115 No External Input or Polled
Data
Informational
Yes
A116 Temperature Overrange
(Petroleum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A117 Density Overrange (Petro-
leum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A118 Discrete Output 1 Fixed
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A119 Discrete Output 2 Fixed
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A120 Curve Fit Failure (Concen-
tration)
Informational
Applies only to transmitters with
the concentration measurement
application.
No
A121 Extrapolation Alarm (Con-
centration)
Informational
Applies only to transmitters with
the concentration measurement
application.
Yes
A131 Meter Verification in Pro-
gress: Outputs to Last
Measured Value
Informational
Applies only to transmitters with
Smart Meter Verification.
Yes
A132 Sensor Simulation Active
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
A141 DDC trigger(s) have com-
pleted
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Yes
5.6 Configure informational parameters
The informational parameters can be used to identify or describe your flowmeter but they
are not used in transmitter processing and are not required.
The informational parameters include:
Device parameters
-
Descriptor
Configure device options and preferences
94 Micro Motion
®
Model 2700 Transmitters with Analog Outputs