6. Lamp & lamp house
6.6 Resetting the lamp parameters
CAUTION: The “LAMP INFO” parameters which are stored on a chip MUST be updated after each installation
of an xenon bulb lamp inside the lamp house. Neglecting this update will result in poor performance and short
life time of the xenon bulb lamp.
How to reset the lamp parameters?
1. Start up the projector but do not ignite the lam p.
2. Go via the Communicator touch panel to the menu “Installation”>“Lamp”>“Lamp information”.
3. While the Lamp information window is displayed, tip on Reset (1).
A reset message is displayed (2).
4. Fill out the article number of the new lamp (3a)
Or,
click on Select (3b) to display a list of possible article numbe rs (4). Select a article number (5) and click OK (6).
The software will check if the entered article number is a valid number (7).
5. Fill out the serial number of the lamp (8).
6. Click Reset (9).
Image 6-21
Reset lamp info
CAUTION: For more information about using the Communicator touch panel consult the user’s guide of the
communicator touch panel.
R59770091 DP-1500 06/01/2009 59