Agilent Technologies E4400-90335 Stereo System User Manual


 
ESG Family Signal Generators Assembly-Level Troubleshooting with Block Diagrams
A12 Synthesizer/Doubler ABUS Nodes (ESG-A & ESG-D Series)
Service Guide 2-37
A12 Synthesizer/Doubler ABUS Nodes (ESG-A & ESG-D
Series)
NOTE The node voltages given in the following table are approximate values based
on a sample of signal generators. Your signal generator may not reflect these
exact values. Additionally, the resolution of these values varies from node to
node. As a guideline, interpret your measurements based on the number of
decimal places shown for the expected voltage.
A12 Synthesizer/Doubler ABUS Nodes
Test Conditions
Node Voltages (Corrected Values in Vdc)
F2
RF_OUT
TUNE
LOOP
10V
FM
PRESET; Freq 500.000001 MHz;
No Modulation
4 0.4 to
0.7
3.0 to 4.8 ≈−0.6 9.9 to
10.1
< 0.2
PRESET; Freq 750 MHz; No Modulation 5.5 0.4 to
0.7
10.2 to
12.8
≈−1.5 9.9 to
10.1
< 0.2
PRESET; Freq 1000 MHz; No Modulation 7.2 0.4 to
0.7
17.7 to
23.2
≈−5.5 9.9 to
10.1
< 0.2
+1 Vdc Applied to EXT 1 INPUT:
PRESET; FM On; FM Source Ext 1 DC
≈−2.0
+1 Vdc Applied to EXT 1 INPUT:
PRESET; Freq < 250 MHz; FM On; FM
Source Ext 1 DC
≈−2.0