Agilent Technologies E4400-90335 Stereo System User Manual


 
ESG Family Signal Generators Assembly Replacement
A2DS1 Fluorescent Lamp
Service Guide 5-9
A2DS1 Fluorescent Lamp
Tools Required
Phillips #0 screwdriver
Removal
1. Remove the display (A2). (Refer to the “A2 Display” replacement procedure on page 5-6.)
2. Loosen the screw (item 1 in Figure 5-4) on the compartment door (item 2) of the display.
Remove the door.
3. Remove the fluorescent lamp (A2DS1) from the compartment.
Replacement
1. Reverse the removal procedure. Ensure the lamp is correctly positioned in the reflective
paper of the compartment.
Figure 5-4 A2DS1 Fluorescent Lamp