Agilent Technologies E4400-90335 Stereo System User Manual


 
Assembly Replacement ESG Family Signal Generators
Instrument Cover
5-64 Service Guide
Instrument Cover
Tools Required
T-15 TORX screwdriver
T-20 TORX screwdriver
Removal
1. Remove the two strap handles (item 1 in Figure 5-39) from each side of the signal
generator by loosening the two screws (item 2) on each handle.
2. Remove the four bottom feet (item 3).
3. Remove the four rear feet (item 4) from the signal generator by removing the four
screws (item 5) that secure them.
4. Slide the instrument cover (item 6) off the back of the signal generator.
Replacement
1. Reverse the removal procedure and do the following:
Torque the rear feet screws (item 4) to 21 in-lbs.
Torque the strap handle screws (item 2) to 21 in-lbs.