Agilent Technologies E4400-90335 Stereo System User Manual


 
Assembly Replacement ESG Family Signal Generators
Rear Panel
5-66 Service Guide
Rear Panel
Tools Required
T-10 TORX screwdriver
3/16-inch wrench or nut driver
9/32-inch wrench or nut driver
5/16-inch open-end wrench
5/8-inch wrench or nut driver
long nose pliers
Removal
1. Remove the instrument cover. (Refer to the “Instrument Cover” replacement procedure
on page 5-64.)
2. Remove the top cover by removing the 11 screws that secure it.
3. Remove the bottom cover by removing the 15 screws that secure it.
4. Disconnect A16W1 from the power supply (A4). (See Figure 5-40.) A16W1 can be
accessed through an opening on the bottom of the power supply shield (item 1).
5. Disconnect all remaining rear panel cables (if any) from the various assemblies within
the signal generator.
NOTE Figure 5-40 represents an Option UN7/UND.