Chapter 4 161
Making Measurements
Using Baseband I/Q Inputs (Option B7C)
Setting Up Baseband I/Q Inputs
Option B7C adds two keys that let you adjust the I/Q inputs; the I/Q
Setup key and the I/Q Range key. Both keys are located under the
Input/Output front panel key in the Input menu.
The
I/Q Range key lets you select one of four levels as an upper limit for
the signal being applied to the Baseband I/Q inputs. The level may be
selected in units of dBm, dBmv, dBuv, Volts, and Watts. The following
table shows the four-level selections available for each unit of measure:
The default is 1 Volt.
The
I/Q Range power levels in Table 4-8 are based on an I/Q Input Z of 50
. I/Q Range voltage levels are independent of I/Q Input Z.
If I/Q Range is set below the default and the error message Input
Overload
is displayed, this value may be adjusted to its maximum.
Beyond that point, the signal must be attenuated to preserve the
measurement accuracy. Using a lower value than the default can
provide an increased dynamic measurement range.
I/Q Setup lets you adjust the following:
I Offset - Use
I Offset to input a value in Volts to offset the measured I
value. The default value is 0.0000 Volts, the min/max values are
±
2.5600 Volts. The tuning increment depends on the I/Q Range setting
as shown in Table 4-9. This value only affects the displayed results,
and does not appear as a correcting voltage at the probe.
Table 4-8 I/Q Range Settings by Displayed Unit of Measure
Unit of
Measure
Highest
Setting
Lowest
Setting
dBm
13.0 dBm 7.0 dBm 1.0 dBm 5.1 dBm
dBmv
60 dBmv 54 dBmv 48 dBmv 41.9 dBmv
dBuv
120.0 dBuv 114.0 dBuv 108 dBuv 101.9 dBuv
Volts
1 Volt 500 mV 250 mV 125 mV
Watts
20.0 mW 5.0 mW 1.2 mW 310.0 uW