Honeywell SMV 3000 Satellite Radio User Manual


 
1/99 SMV 3000 Transmitter Users Manual 63
6.8 TempConf Configuration - PV3, Continued
Sensor Type
Identify and select the type of sensor that is connected to the transmitter as
its input for process temperature PV3. This will set the appropriate LRL
and URL data in the transmitter automatically.
Table 15 shows the pre-programmed temperature sensor types and the
rated measurement range limits for a given sensor selection.
Table 15 Sensor Types for PV3 Process Temperature Input
Sensor Type Rated Temperature Range Limits
°C °F
PT100 D
d -200 to 450 -328 to 842
Type E 0 to 1000 32 to 1832
Type J 0 to 1200 32 to 2192
Type K -100 to 1250 -148 to 2282
Type T -100 to 400 -148 to 752
d
Factory setting.
ATTENTION
Whenever you connect a different sensor as the transmitters input, you
must also change the sensor type configuration to agree. Otherwise, range
setting errors may result.