Honeywell SMV 3000 Satellite Radio User Manual


 
1/99 SMV 3000 Transmitter Users Manual 109
9.5 Replacing Meter Body Center Section, Continued
Procedure, continued
Table 29 Replacing Meter Body Center Section, Continued
Step Action
4
Remove nuts from bolts that hold process heads to center section.
Remove process heads and bolts
5
Remove O-ring and clean interior of process head using soft bristle
brush and suitable solvent.
6
Replace O-ring.
7
Coat threads on process head bolts with anti-seize compound such
as Neverseize or equivalent.
8
Carefully assemble process heads and bolts to new center section.
Finger tighten nuts.
22370
Nuts
O-ring
O-ring
Bolts
Process
head
Center
section
Process
head
Flex-Tape Assembly
Continued on next page