134 SMV 3000 Transmitter User’s Manual 1/99
11.4 Diagnostic Messages, Continued
Diagnostic Messages,
continued
Table 34 Informational Status Message Table
SMV Status SCT Status Message SFC Display Message TDC Status Message Possible Cause What to Do
6-3 2 Wire TC PV3
STATUS TAG ID.
2 WIRE TC PV3
2 WIRE TC PV3
PV3 input is being provided by
2-wire Thermocouple (T/C) type.
Nothing – Information only. However,
this may indicate a problem if sensor
type does not match the sensor
physically connected to transmitter.
6-0 2 Wire RTD PV3
STATUS TAG ID.
2 WIRE RTD PV3
2 WIRE RTD PV3
PV3 input is being provided by
2-wire RTD type.
Nothing – Information only. However,
this may indicate a problem if number
of wires displayed does not match
number of RTD leads physically
connected to transmitter; or if sensor
type should be thermocouple.
6-1 3 Wire RTD PV3
STATUS TAG ID.
3 WIRE RTD PV3
3 WIRE RTD PV3
PV3 input is being provided by
3-wire RTD type.
Nothing – Information only. However,
this may indicate a problem if number
of wires displayed does not match
number of RTD leads physically
connected to transmitter; or if sensor
type should be thermocouple.
6-2 4 Wire RTD PV3
STATUS TAG ID.
4 WIRE RTD PV3
4 WIRE RTD PV3
PV3 input is being provided by
4-wire RTD type.
Nothing – Information only. However,
this may indicate a problem if number
of wires displayed does not match
number of RTD leads physically
connected to transmitter; or if sensor
type should be thermocouple.
4-3 PV2 Sensor = AP
- STATUS 4-3
Sensor type for the current SMV
is absolute pressure.
Nothing – Information only.
4-4 PV2 Sensor = GP
- STATUS 4-4
Sensor type for the current SMV
is gauge pressure.
Nothing – Information only.
- Write Protected
URV 1 . TAG ID.
WRITE PROTECTED
-
The value could not be written
because the transmitter is write
protected.
The hardware jumper within the
device must be repositioned in order
to permit write operations.