33
Getting Started
Setting up N9061A on the X-Series Analyzer
Logging
The N9061A application allows the logging of errors. These errors comprise details of command errors
and legacy commands that have been received but are not supported by the N9061A application. To
enable and view the error log select the
Mode Setup hardkey from the front panel. Then select the
Logging softkey.
Figure 1-3 Logging Menu Map
Previous Page/Next Page
When you are in the Logging menu, the main Signal Analysis display is obscured by the logging page.
The most recent log starts from the bottom of the window.
Previous Page and Next Page allow you to
scroll through the log file. To include commands sent to the analyzer since the log window display was
opened, press
Refresh.
Cmd Error Log
The
Cmd Error Log option allows you to turn the command error logging on or off. The default is Off.
Logging should not be used in a secure environment. When set to
On, all error messages are stored in a
log file, regardless of whether they have been displayed on the screen. When set to
Off, no further
command error messages are written to the log file. Switching
Cmd Error Log to Off does not clear the
log file.
The log file is also stored as a text file, called Logfile.txt, on the instrument. It is stored in the D:\
drive, in a folder called \User_My_Documents\[USERNAME]\My Documents\RLC\data.
The maximum size of the log is 10 MB. When the file reaches its maximum size, the first ten percent of
the file is automatically discarded, to clear space for subsequent error messages.
Previous Page
Logging
Logging
Next Page
Clear Log
Refresh
Cmd Error Log
On Off