Agilent Technologies N9030a Stereo System User Manual


 
152
Programming Commands
ERR Error
ERR
Error
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
The ERR command returns a list of three-digit error codes if errors are present. A code of “0” means that
there are no errors present. Executing ERR? clears all GPIB errors.
If a command is a valid legacy command but not accepted by the N9061A application, no error message
is generated and the response to ERR? is 0. However, if logging is enabled, the N9061A application
command log registers a “Cmd not Supported” error.
If a command is not a valid legacy command, a command error is generated; CMD ERR is displayed on
the front panel and the response to ERR? is 112. If logging is enabled then Cmd Error is written to the
command error log.
Error codes are provided in RLC mode for some X-series errors such as external reference, hardware and
alignment errors. The X-series error codes are translated to 8560 series error codes so that an error query
returns the legacy instrument error code. To review the error via the front panel, select the
System
hardkey and then select
Show > Errors. The following table shows the X-series error codes and the
translated value.
Table 4-6
Error Code Description
112 CMD ERR
ERR ?
LF
CR
SP
,
;