Programming Commands
MEANPWR Mean Power measurement
MEANPWR
Mean Power measurement
Syntax
Legacy Products
8560 series
Description
The MEANPWR command measures the average power of the carrier during that portion of the time
when it is on. The on state is defined as the time when the signal is within a selected number of dB of its
peak level. The range of amplitudes that is defined as the on state can be set with the command. The
amplitude range is set relative to the peak value of the signal.
NOTE The MEANPWR command is similar to the CARRON command (page 123),
except that the CARRON command defines ‘on’ as that time when the signal is
within 20 dB of its peak level.
MEANPWR TRA
S
P
TRB
, number ,
?
LF
CR
SP
,
;