Programming Commands
MKD Marker Delta
MKD
Marker Delta
Syntax
Step Increment: by 1/10 of the frequency span
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
The MKD command computes the frequency and amplitude difference of the active marker and the delta
marker. These values are displayed on the screen.
If a delta marker is not displayed on the screen, the MKD command places one at the specified frequency
or on the left or right hand edge of the display. If an active marker is not displayed on the screen, the
MKD command places an active marker at the center of the screen.
NOTE For 8566A/B and 8568A/B, the functions of the MKD command are identical to
the M3 command (page 229).