Agilent Technologies N9030a Stereo System User Manual


 
176
Programming Commands
I2 [two] Set RF Coupling to AC
NOTE The X-Series analyzers only have a single RF input port.
Option 526 9 kHz 26.5 GHz 10 MHz 26.5 GHz
Table 4-20 MXA Series Analyzer Frequency Coupling Specifications
DC Coupled Range AC Coupled Range
Analyzer Model
(N9020A)
Min. Freq. Max. Freq. Min. Freq. Max. Freq.
Option 503 20 Hz 3.6 GHz 10 MHz 3.6 GHz
Option 508 20 Hz 8.4 GHz 10 MHz 8.4 GHz
Option 513 20 Hz 13.6 GHz 10 MHz 13.6 GHz
Option 526 20 Hz 26.5 GHz 10 MHz 26.5 GHz
Table 4-21 PXA Series Analyzer Frequency Coupling Specifications
DC Coupled Range AC Coupled Range
Analyzer Model
(N9030A)
Min. Freq. Max. Freq. Min. Freq. Max. Freq.
Option 503 3 Hz 3.6 GHz 10 MHz 3.6 GHz
Option 508 3 Hz 8.4 GHz 10 MHz 8.4 GHz
Option 513 3 Hz 13.6 GHz 10 MHz 13.6 GHz
Option 526 3 Hz 26.5 GHz 10 MHz 26.5 GHz
Table 4-19 EXA Series Analyzer Frequency Coupling Specifications
DC Coupled Range AC Coupled Range
Analyzer Model
(N9010A)
Min. Freq. Max. Freq. Min. Freq. Max. Freq.