Programming Commands
MKPX Marker Peak Excursion
MKPX
Marker Peak Excursion
Syntax
Preset State: 6 dB
Step Increment: 1 dB
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Specifies the minimum signal excursion for the analyzer’s internal peak identification routine.
The default value is 6 dB. In this case, any signal with an excursion of less than 6 dB on either side of the
marker would not be identified. Thus, if an MKPK NH command were to be executed on such a signal,
the analyzer would not place a marker on this signal peak.
MKPX
?
;S
P
real
S
P
UP
DN
OA
DB
LF
CR
SP
,