Agilent Technologies N9030a Stereo System User Manual


 
212
Programming Commands
KSx External Trigger
KSx
External Trigger
Syntax
Legacy Products
8566A/B, 8568A/B
Description
The KSx command activates the normal external trigger mode. When the KSx command is executed, the
RF input signal is only displayed when the external trigger level exceeds the trigger threshold level.
NOTE If an 8566A/B or an 8568A/B analyzer is in zero span and the sweep time is less
than 20 msec, the display only gets refreshed when a fresh trace has been taken.
This can cause the displayed trace to flicker.
In the N9061A application of X-Series analyzers, all traces are displayed
continuously, so all traces are therefore free of flicker.
NOTE The functions of the command KSx are identical to the TM EXT command
(page 337).
KSx
;