Agilent Technologies N9030a Stereo System User Manual


 
153
Programming Commands
ERR Error
Table 4-7
X-series
error code
Description 8560 series
error code
Description
40 TG Alignment Failure 758 SYSTEM: Unknown system error
42 RF Alignment Failure 758 SYSTEM: Unknown system error
44 IF Alignment Failure 758 SYSTEM: Unknown system error
46 LO Alignment Failure 758 SYSTEM: Unknown system error
48 ADC Alignment Failure 758 SYSTEM: Unknown system error
50 FM Demod Alignment
Failure
758 SYSTEM: Unknown system error
54 Extended Align Failure
Sum
758 SYSTEM: Unknown system error
71 Characterize Preselector
Failure
758 SYSTEM: Unknown system error
-200.3310 Execution Error;
Preselector Centering
failed
758 SYSTEM: Unknown system error
503 Frequency Reference
Unlocked
336 10 MHz Ref Cal oscillator failed to lock
when going to internal 10 MHz
reference.
505 2nd LO Unlocked 336 10 MHz Ref Cal oscillator failed to lock
when going to internal 10 MHz
reference.
509 LO Unlocked 300 YTO UNL: YTO (1st LO) phase-locked
loop (PLL) is unlocked.
513 IF Synthesizer Unlocked 450 IF SYSTM: IF hardware failure. Check
other error messages.
515 Calibration Oscillator
Unlocked
336 10 MHz Ref: Cal oscillator failed to lock
when going to internal 10 MHz
reference
521 External Ref missing or
out of range
905,333 EXT REF: Unable to lock cal oscillator
when set to external reference. Check
that the external reference is within
tolerance.
600 UNLK: 600 MHz reference
oscillator PLL is unlocked