96 Chapter 4
Option P03 - Preamplifier
Other Preamp Specifications
Description Specifications Supplemental Information
Frequency Response − Preamp On
(Option P03)
Refer to the footnote for
Band Overlaps on page 15.
Maximum error relative to
reference condition (50 MHz)
Input attenuation 0 dB
Swept operation
a
a. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An
additional error source, the error in switching between swept and FFT sweep types, is nominally ±0.01
dB and is included within the “Absolute Amplitude Error” specifications.
20 to 30 °C 5 to 50 °C
95
th
Percentile (≈2σ)
20 to 30 °C
100 kHz to 3.6 GHz
b
b. Electronic attenuator (Option EA3) may not be used with preamp on.
±0.28 dB (nominal)