Agilent Technologies N9010A Stereo System User Manual


 
Chapter 9 123
W-CDMA Measurement Application
Amplitude
Description Specifications Supplemental Information
Adjacent Channel Power
(ACPR; ACLR)
Single Carrier
Minimum power at RF Input 36 dBm (nominal)
ACPR Accuracy
a
a. The accuracy of the Adjacent Channel Power Ratio will depend on the mixer drive level and whether
the distortion products from the analyzer are coherent with those in the UUT. These specifications
apply even in the worst case condition of coherent analyzer and UUT distortion products. For ACPR
levels other than those in this specifications table, the optimum mixer drive level for accuracy is
approximately 37 dBm (ACPR/3), where the ACPR is given in (negative) decibels.
RRC weighted, 3.84 MHz noise
bandwidth, method = IBW or Fast
b
Radio Offset Freq
MS (UE) 5 MHz ±0.22 dB At ACPR range of 30 to 36 dBc with
optimum mixer level
c
MS (UE) 10 MHz ±0.34 dB At ACPR range of 40 to 46 dBc with
optimum mixer level
d
BTS 5 MHz ±1.07 dB At ACPR range of 42 to 48 dBc with
optimum mixer level
e
BTS 10 MHz ±1.00 dB At ACPR range of 47 to 53 dBc with
optimum mixer level
d
BTS 5 MHz ±0.44 dB
At 48 dBc non-coherent ACPR
d
Dynamic Range RRC weighted, 3.84 MHz noise
bandwidth
Noise
Correction
Offset Freq Method Dynamic Range
(typical)
f
Optimum ML
(nominal)
off 5 MHz IBW 68 dB 8 dBm
off 5 MHz Fast 67 dB 9 dBm
off 10 MHz IBW 74 dB 2 dBm
on 5 MHz IBW 73 dB 8 dBm
on 10 MHz IBW 76 dB 2 dBm
RRC Weighting Accuracy
g
White noise in Adjacent Channel 0.00 dB (nominal)
TOI-induced spectrum 0.001 dB (nominal)
rms CW error 0.012 dB (nominal)