Agilent Technologies N9010A Stereo System User Manual


 
Chapter 20 221
DVB-T/H Measurement Application
N6153A, DVB-T/H Measurements Application
Description Specifications Supplemental Information
64 QAM EVM
ML
a
= –20 dBm
20 to 30 °C
FFT Size = 2048, GuardInterval =
1/32,
alpha = 1
EVM
Operating range 0 to 8%
Floor 0.64% (blind equalizer ON)
0.73% (blind equalizer OFF)
Accuracy
from 0.7% to 1.2%
from 1.2% to 2.0%
from 2.0% to 8.0%
±0.30%
±0.20%
±0.20%
MER
Operating range 22.00 dB
Floor 43.88 dB (blind equalizer ON)
42.73 dB (blind equalizer OFF)
Accuracy
from 38 to 43 dB
from 34 to 38 dB
from 22 to 34 dB
±2.62 dB
±1.02 dB
±0.48 dB
Frequency Error
b
Range –100 kHz to 100 kHz
Accuracy
±10 Hz+tfa
c
Phase Jitter
Range 0 to 0.0349 rad
Resolution 0.0001 rad
Quad Error
Range –4 deg to 5 deg
Accuracy ±0.090 deg
Amplitude Imbalance
Range –5% to +5%