Agilent Technologies N9010A Stereo System User Manual


 
Chapter 1 29
Agilent EXA Signal Analyzer
Amplitude Accuracy and Range
Amplitude Accuracy and Range
Description Specifications Supplemental
Information
Measurement Range Displayed Average Noise Level to +23 dBm
Preamp On Displayed Average Noise Level to +23 dBm Option P03
Input Attenuation Range 0 to 60 dB, in 10 dB steps Standard
Input Attenuation Range 0 to 60 dB, in 2 dB steps With Option FSA
Description Specifications Supplemental Information
Maximum Safe Input Level Applies with or without preamp
(Option P03)
Average Total Power +30 dBm (1 W)
Peak Pulse Power
<10 μs pulse width,
<1% duty cycle
input attenuation 30 dB
+50 dBm (100 W)
DC volts
DC Coupled ±0.2 Vdc
AC Coupled ±70 Vdc
Description Specifications Supplemental Information
Display Range
Log Scale Ten divisions displayed;
0.1 to 1.0 dB/division in 0.1 dB steps, and
1 to 20 dB/division in 1 dB steps
Linear Scale Ten divisions