172 Chapter 14
1xEV-DO Measurement Application
Measurements
Measurements
Description Specifications Supplemental Information
Channel Power
1.23 MHz Integration BW
Input signal must not be bursted
Minimum power at RF input −50 dBm (nominal)
Absolute power accuracy
a
20 to 30 °C
a. Absolute power accuracy includes all error sources for in-band signals except mismatch errors and
repeatability due to incomplete averaging. It applies when the mixer level is high enough that measure-
ment floor contribution is negligible.
±0.94 dB
±0.27 dB (typical)
Measurement floor −84 dBm (nominal)