Agilent Technologies N9010A Stereo System User Manual


 
Chapter 1 17
Agilent EXA Signal Analyzer
Frequency and Time
Achievable Initial Calibration
Accuracy
±1.4 × 10
6
Settability
±2 × 10
8
Residual FM
Center Frequency = 1 GHz
10 Hz RBW, 10 Hz VBW
10 Hz × N p-p in 20 ms
c
,
nominal
a. Calibration accuracy depends on how accurately the frequency standard was adjusted to 10
MHz. If the adjustment procedure is followed, the calibration accuracy is given by the specifi-
cation “Achievable Initial Calibration Accuracy.”
b. For periods of one year or more.
c. N is the LO multiplication factor.
Description Specifications Supplemental Information