Agilent Technologies N9010A Stereo System User Manual


 
Chapter 4 95
Option P03 - Preamplifier
Other Preamp Specifications
Description Specifications Supplemental Information
Displayed Average Noise Level
(DANL) Preamp On
(Option P03)
a
a. DANL for zero span and swept is normalized in two ways and for two reasons. DANL is measured in a
1 kHz RBW and normalized to the narrowest available RBW, because the noise figure does not depend
on RBW and 1 kHz measurements are faster. The second normalization is that DANL is measured with
10 dB input attenuation and normalized to the 0 dB input attenuation case, because that makes DANL
and third order intermodulation test conditions congruent, allowing accurate dynamic range estimation
for the analyzer.
Input terminated,
Sample or Average detector
Averaging type = Log
0 dB input attenuation
IF Gain = Any setting
Refer to the footnote for
Band Overlaps on page 15.
1 Hz Resolution Bandwidth
Preamp On
20 to 30 °C 5 to 50 °CTypical Nominal
Option P03
100 kHz to 1 MHz
b
b. Specifications apply only when the Phase Noise Optimization control is set to “Best Phase Noise at off-
set > 30 kHz.”
146 dBm
1 MHz to 10 MHz 161 dBm
10 MHz to 2.1 GHz 161 dBm 159 dBm 163 dBm
2.1 GHz to 3.6 GHz 160 dBm 158 dBm 162 dBm