Agilent Technologies N9010A Stereo System User Manual


 
Chapter 1 49
Agilent EXA Signal Analyzer
Dynamic Range
400 MHz to 1.7 GHz +10 dBm 80 dBc
1.7 to 3.6 GHz +12 dBm 84 dBc
3.6 to 5.1 GHz +10 dBm 80 dBc
5.1 to 7 GHz +12 dBm 86 dBc
7 to 13.6 GHz +10 dBm 80 dBc
13.6 to 26.5 GHz +7 dBm 74 dBc
a. See the IF Prefilter Bandwidth table in the Gain Compression specifications on page 42.
When the tone separation condition is met, the effect on TOI of the setting of IF Gain is negli-
gible. TOI is verified with IF Gain set to its best case condition, which is IF Gain = Low.
b. TOI is verified with two tones, each at 18 dBm at the mixer, spaced by 100 kHz.
c. TOI = third order intercept. The TOI is given by the mixer tone level (in dBm) minus (distor-
tion/2) where distortion is the relative level of the distortion tones in dBc.
d. The distortion shown is computed from the warranted intercept specifications, based on two
tones at 30 dBm each, instead of being measured directly. The choice of 30 dBm is based
on historic and continuing industry practice.
Description Specifications Supplemental Information