Agilent Technologies N9010A Stereo System User Manual


 
42 Chapter 1
Agilent EXA Signal Analyzer
Dynamic Range
Dynamic Range
Gain Compression
Description Specifications Supplemental Information
1 dB Gain Compression Point
(Two-tone)
abc
a. Large signals, even at frequencies not shown on the screen, can cause the analyzer to incor-
rectly measure on-screen signals because of two-tone gain compression. This specification
tells how large an interfering signal must be in order to cause a 1 dB change in an on-screen
signal.
b. Specified at 1 kHz RBW with 100 kHz tone spacing. The compression point will nominally
equal the specification for tone spacing greater than 5 times the prefilter bandwidth. At
smaller spacings, ADC clipping may occur at a level lower than the 1 dB compression point.
Maximum power at
mixer
d
(nominal)
20 MHz to 26.5 GHz +9 dBm (nominal)
Clipping (ADC Over Range)
Any signal offset 10 dBm
Low frequency exceptions
d
IF Gain set to Low
Signal offset >5 times IF prefilter bandwidth
+12 dBm (nominal)
IF Prefilter Bandwidth
Zero Span or Swept: Sweep Type = FFT:
RBW FFT Width 3 dB Bandwidth, nominal
3.9 kHz < 4.01 kHz 8.9 kHz
4.3 - 27 kHz < 28.81 kHz 79 kHz
30 - 160 kHz < 167.4 kHz 303 kHz
180 - 390 kHz < 411.9 kHz 966 kHz
430 kHz - 8 MHz < 7.99 MHz 10.9 MHz