Chapter 25 261
Option EMC Precompliance Measurements
Amplitude
Amplitude
Description Specifications Supplemental Information
EMI Average Detector Used for CISPR-compliant average
measurements and, with 1 MHz
RBW, for frequencies above 1 GHz
Default Average Type All filtering is done on the linear
(voltage) scale even when the display
scale is log.
Quasi-Peak Detector Used with CISPR-compliant RBWs,
for frequencies 1 GHz
Absolute Amplitude Accuracy for
reference spectral intensities
Meets CISPR standards
a
a. CISPR 16-1-1 (2007)
Relative amplitude accuracy versus
pulse repetition rate
Meets CISPR standards
a
Quasi-Peak to average response ratio
Meets CISPR standards
a
Dynamic range
Pulse repetition rates 20 Hz
Meets CISPR standards
a
Pulse repetition rates 10 Hz Does not meet CISPR standards in
some cases with DC pulse excitation.
RMS Average Detector
Meets CISPR standards
a