Agilent Technologies N9010A Stereo System User Manual


 
Chapter 17 201
Single Acquisition Combined Fixed WiMAX Measurement Application
N9074A-XFP, Single Acquisition Combined Fixed WiMAX Measurements
d. The relative accuracy is a measure of the ratio of the power at the offset to the main channel power. It
applies for spectrum emission levels in the offsets that are well above the dynamic range limitation.
Description Specifications Supplemental Information
64QAM EVM
ML
a
= –10 dBm
20 to 30 °C
a. ML (mixer level) is RF input power minus attenuation
10 MHz bandwidth profile.
Code Rate: 3/4
EQ Seq
Track Phase On
Track Amp Off
Track Timing Off
EVM
Operating range 0.1 to 8% (nominal)
Floor –45.0 dB (0.57%) (nominal)
Accuracy
b
b. The accuracy specification applies when the EVM to be measured is well above the measurement
floor. When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy
errors. The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors
depend on the EVM of the UUT and the floor as follows: error = sqrt(EVMUUT
2
+ EVMsa
2
) -
EVMUUT, where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the
analyzer in percent.
from 0.5% to 2.0% ±0.30% (nominal)
from 2.0% to 8.0% ±0.10% (nominal)
I/Q Origin Offset
UUT Maximum Offset –10 dBc (nominal)
Analyzer Noise Floor –50 dBc (nominal)
Frequency
Range ±100 kHz (nominal)
Accuracy
±10 Hz+tfa
c
c. tfa = transmitter frequency frequency reference accuracy.