Agilent Technologies N9010A Stereo System User Manual


 
Chapter 18 205
Single Acquisition Combined WLAN Measurement Application
N9077A, Combined WLAN 802.11a or 802.11g-OFDM Measurements
Absolute
20 – 30°C ±1.58 dB
30 MHz offset
Dynamic Range, relative 67.5 dB (nominal)
Sensitivity, absolute –80.7 dBm (nominal)
Accuracy
Relative ±0.66 dB
Absolute
20 – 30°C ±1.60 dB
a. The dynamic range specification is the ratio of the channel power to the power in the offset specified.
The dynamic range depends on the measurement settings, such as peak power or integrated power.
Dynamic range specifications are based on default measurement settings, with detector set to average,
and depend on the mixer level. Default measurement settings include 100 kHz RBW.
b. This dynamic range specification applies for the optimum mixer level, which is about –12.89 dBm.
Mixer level is defined to be the average input power minus the input attenuation.
c. The sensitivity is specified with 0 dB input attenuation. It represents the noise limitations of the ana-
lyzer. It is tested without an input signal. The sensitivity at this offset is specified in the default 100 kHz
RBW, at a center frequency of 1 GHz.
d. The relative accuracy is a measure of the ratio of the power at the offset to the main channel power. It
applies for spectrum emission levels in the offsets that are well above the dynamic range limitation.
Description Specifications Supplemental Information
64QAM EVM
ML
a
= –10 dBm
20 to 30 °C
Code Rate:3/4
EQ Seq
Track Phase On
Track Amp Off
Track Timing Off
EVM
Operating range 0.1 to 8% (nominal)
Floor –42 dB (0.77%) (nominal)
Accuracy
b
from 0.5% to 2.0% ±0.50% (nominal)
from 2.0% to 8.0% ±0.30% (nominal)
Description Specifications Supplemental Information