Agilent Technologies N9010A Stereo System User Manual


 
226 Chapter 21
ISDB-T Measurement Application
N6155A, ISDB-T/TSB Measurement Application
Description Specifications Supplemental Information
Spectrum Emission Mask Limit Type
Manual
JEITA (ARIB-B31) according to
P 0.025 W;
0.025 W < P 0.25 W;
0.25 W < P 2.5 W;
P > 2.5 W
(P is the channel power)
ABNT Non-Critical
ABNT Sub-Critical
ABNT Critical
•ISDB-T
SB
5.60 MHz Integration BW
RBW = 10.0 kHz
3.0 MHz offset
Dynamic Range, relative
ab
a. The dynamic range specification is the ratio of the channel power to the power in the offset specified.
The dynamic range depends on the measurement settings, such as peak power or integrated power.
Dynamic range specifications are based on default measurement settings, with detector set to average,
and depend on the mixer level. Default measurement settings include 10.0 kHz RBW.
b. This dynamic range specification applies for the optimum mixer level, which is about –16 dBm. Mixer
level is defined to be the average input power minus the input attenuation.
82.5 dB 89.7 dB (typical)
Sensitivity, absolute
c
–101.5 dBm –107.5 dBm (typical)
Accuracy
Relative
d
±0.61 dB
Absolute
20 – 30°C ±1.05 dB
4.36 MHz offset
Dynamic Range, relative 83.0 dB 90.1 dB (typical)
Sensitivity, absolute –101.5 dBm –107.5 dBm (typical)
Accuracy
Relative ±0.63 dB
Absolute
20 – 30°C ±1.05 dB