48 Chapter 1
Agilent EXA Signal Analyzer
Dynamic Range
Second Harmonic Distortion
Third Order Intermodulation Distortion
Description Specifications Supplemental
Information
Second Harmonic Distortion
Mixer Level
a
a. Mixer level = Input Level Input Attenuation
SHI
b
(nominal)
b. SHI = second harmonic intercept. The SHI is given by the mixer power in dBm minus the sec-
ond harmonic distortion level relative to the mixer tone in dBc.
Source Frequency
10 MHz to 1.8 GHz 15 dBm +45 dBm
1.75 to 7 GHz 15 dBm +65 dBm
7 GHz to 11 GHz 15 dBm +55 dBm
11 to 13.25 GHz 15 dBm +50 dBm
Description Specifications Supplemental Information
Third Order
Intermodulation Distortion
Tone separation > 5 times IF
Prefilter Bandwidth
a
Verification conditions
b
Refer to the footnote for
Band Overlaps on page 15.
20 to 30°C
Intercept
c
Extrapolated
Distortion
d
Intercept (typical)
100 to 400 MHz +10 dBm 80 dBc +14 dBm
400 MHz to 1.7 GHz +11 dBm 82 dBc +15 dBm
1.7 to 3.6 GHz +13 dBm 86 dBc +17 dBm
3.6 to 5.1 GHz +11 dBm 82 dBc +17 dBm
5.1 to 7 GHz +13 dBm 86 dBc +17 dBm
7 to 13.6 GHz +11 dBm 82 dBc +15 dBm
13.6 to 26.5 GHz +9 dBm 78 dBc +14 dBm
5 to 50°C
10 to 100 MHz
100 to 400 MHz +9 dBm 78 dBc