Agilent Technologies N9010A Stereo System User Manual


 
Chapter 3 89
Option EA3 - Electronic Attenuator, 3.6 GHz
Other Electronic Attenuator Specifications
Description Specifications Supplemental Information
Frequency Response
Maximum error relative to
reference condition (50 MHz)
20 to 30 °C 5 to 50 °C
95
th
Percentile (2σ)
Attenuation = 4 to 24 dB, even
steps
9 kHz to 10 MHz ±0.75 dB ±0.90 dB ±0.32 dB
10 MHz to 50 MHz ±0.65 dB ±0.69 dB ±0.27 dB
50 MHz to 2.2 GHz ±0.48 dB ±0.60 dB ±0.19 dB
2.2 GHz to 3.6 GHz ±0.55 dB ±0.67 dB ±0.20 dB
Attenuation = 0, 1, 2 and odd steps,
3 to 23 dB
10 MHz to 3.6 GHz ±0.30 dB
Description Specifications Supplemental Information
Electronic Attenuator Switching
Uncertainty
E r r or re l a t ive t o r e fer e n c e
condition (50 MHz, 10 dB
mechanical attenuation, 10 dB
electronic attenuation)
Attenuation = 0 to 24 dB
9 kHz to 3.6 GHz
See note
a
a. The specification is ±0.14 dB. Note that this small relative uncertainty does not apply in esti-
mating absolute amplitude accuracy. It is included within the absolute amplitude accuracy for
measurements done with the electronic attenuator. (Measurements made without the elec-
tronic attenuator are treated differently; the absolute amplitude accuracy specification for
these measurements does not include attenuator switching uncertainty.)