Agilent Technologies N9010A Stereo System User Manual


 
Chapter 14 175
1xEV-DO Measurement Application
Measurements
Description Specifications Supplemental Information
Spurious Emissions
Dynamic Range, relative 91.9 dB 97.1 dB (typical)
Sensitivity, absolute 79.4 dBm 85.4 dBm (typical)
Accuracy, absolute
20 Hz to 3.6 GHz ±0.38 dB (95% confidence)
3.5 GHz to 8.4 GHz ±1.22 dB (95% confidence)
8.3 GHz to 13.6 GHz ±1.59 dB (95% confidence)
Description Specifications Supplemental Information
QPSK EVM
25 dBm ML
a
15 dBm
20 to 30 °C
a. ML (mixer level) is RF input power minus attenuation
RF input power range is accordingly
determined to meet Mixer level.
EVM
Operating range 0 to 25%
Floor 1.5%
Accuracy
b
b. The accuracy specification applies when the EVM to be measured is well above the measurement floor.
When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy errors.
The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors depend
on the EVM of the UUT and the floor as follows: error = sqrt(EVMUUT
2
+ EVMsa
2
) – EVMUUT,
where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the analyzer in
percent.
±1.0%
I/Q origin offset
DUT Maximum Offset 10 dBc (nominal)
Analyzer Noise Floor 50 dBc (nominal)
Frequency Error
Range
±30 kHz (nominal)
Accuracy
±5 Hz + tfa
c
c. tfa = transmitter frequency x frequency reference accuracy.