Barco MGP D5 Home Theater System User Manual


 
Using MediCal Pro
78 MGP D5 Dicom Theater Lite
Using MediCal Pro
After installation of the hardware and software, you must start
MediCal Pro to calibrate the display(s).
Proceed as follows:
1. Start MediCal Pro. If appropriate, you can connect to MediCal
Administrator.
2. During startup, MediCal Pro will detect the connected review
display(s). They will be added automatically to the system
configuration in MediCal Pro.
Note: In case the review displays were not automatically
added to the configuration, you can add them manually, by
running the Configuration Setup Wizard.
3. For all the devices in the configuration, check if the DPMS
setting is turned on. Therefore, right-click on the display icon
and select Properties... from the drop-down menu. Then click
on Details...
Check if the Powersave option is checked. If not, check it.
This is necessary to use the DPMS possibilities of the imaging
board.
4. For all the devices in the configuration, define (if necessary)
and select a Preset.
5. After selecting the Preset, MediCal Pro starts consistency
calibration automatically.
Please refer to the MediCal Pro User Guide for more information.