Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
74
Press the [DEFINE CONSTANT] softkey (F2).
Press the [DEFINE K1] softkey (F1).
Enter 386.4 and press the [ENTER] softkey (F1).
Press the [Analys] hardkey.
Press the [DEFINE FUNCTION] softkey (F1).
Press the [DEFINE F1] softkey (F1).
Press the [CONSTANT (K1-K5)] softkey (F3).
Press the [CONSTANT K1] softkey (F1).
Press the [ * ] softkey (F3).
Press the [OPERATION] softkey (F5).
Press the [MORE] softkey (F9).
Press the [ /jOMEGA( ] softkey (F4).
Press the [OPERATION] softkey (F5).
Press the [MORE] softkey (F9).
Press the [ /jOMEGA( ] softkey (F4).
Press the [MEAS DATA] softkey (F1).
Press the [FREQ RESP 2 / 1] softkey (F5).
Press the [ ) ] softkey (F5).
Press the [ ) ] softkey (F5).
Press the [ENTER] softkey (F7).
Now display the function F1 on Trace B
Press the [Active Trace] hardkey.
Press the [B] softkey (F2).
Press the [Meas Data] hardkey.
Press the [MORE CHOICES] softkey (F9).
Press the [MATH FUNCTION] softkey (F7).
Press the [PWR SPEC CHANNEL 1] softkey (F3)
Press the [F1] softkey (F1).
Trace B will now convert “g/lb.” into “inch/lb.” This same technique could also be
used to convert “g” into velocity terms, or to convert between metric and English
units.