Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
27
Press the [TRIGGER SETUP] softkey (F6)
Press the [ALL CHANNELS] softkey (F8)
Press the [CHANNEL * DELAY] softkey (F9)
Enter -.01 then press the [S] softkey (F1)
Now to check that the trigger parameters are correct,
Press the yellow [Start] hardkey.
The analyzer should show in highlighted text above the trace:
WAITING FOR CH 1 TRIGGER
Tap the hammer on the test structure to see if the measurement will trigger. If it will
not, adjust the trigger level and slope until successful.
Setting Up Time Displays
View both the time trace and the windowed time trace for both channels using the
QUAD display format as follows:
Press the [Disp Format] hardkey.
Press the [QUAD] softkey (F6).
Press the [Active Trace] hardkey.
Press the [A] softkey (F1)
Press the [Meas Data] hardkey.
Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
Press the [TIME CHANNEL 1] softkey (F5)
Press the [Active Trace] hardkey.
Press the [B] softkey (F1)
Press the [TIME CHANNEL 2] softkey (F5)
Press the [Active Trace] hardkey.
Press the [C] softkey (F1)
Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
Press the [MORE CHOICES] softkey (F9)
Press the [WINDOWED TIME CH 1] softkey (F5)
Press the [Active Trace] hardkey.
Press the [D] softkey (F1)
Press the [WINDOWED TIME CH 2] softkey (F5)
Press the [Active Trace] hardkey.
Press the [ABCD] softkey (F7).
Press the [Scale] hardkey.
Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.