Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
52
Choose measurement parameters
Press the [Freq] hardkey
Enter 800 and then press the [Hz] softkey (F2)
Press the [Avg] hardkey.
Press the [NUMBER AVERAGES] softkey (F2), enter 4, then press the
[ENTER] softkey (F1).
Turn on averaging by pressing the [AVERAGE ON OFF] softkey (F1)
until ON is highlighted.
Press the [Window] hardkey
Press the [UNIFORM] softkey (F3).
Set up display parameters
Press the [Disp Format] hardkey.
Press the [UPPER/LOWER] softkey (F3)
Press the [Active Trace] hardkey.
Press the [AB] softkey (F5).
Press the [Meas Data] hardkey.
Press the [FREQ RESP 2 / 1 ] softkey (F6)
Press the [Scale] hardkey.
Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Press the [Active Trace] hardkey.
Press the [B] softkey (F2).
Press the [Trace Coord] hardkey.
Press the [PHASE] softkey (F4).
Set up the AGILENT 35670A source (if amplifier has level control)
Check the following connections:
AGILENT 35670A source connected to amplifier input
amplifier plugged in
amplifier output connected to shaker
force transducer connected to Channel 1
Reduce amplifier level control the lowest setting
Turn on amplifier
Press the [Source] hardkey.
Press the [RANDOM NOISE] softkey (F5)
Press the [LEVEL] softkey (F2)
Enter 0.5, then press the [Vpk] softkey (F1).
Toggle the [SOURCE ON OFF] softkey (F1) until ON is highlighted.