Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
36
Press the [MORE CHOICES] softkey (F9).
Press the [DATA REGISTER] softkey (F8).
Press the [D2] softkey (F2).
Recalling the Spectra from Disk
Use data registers D1 and D2 to contain the spectra used for comparison. D1 is
compared to Channel 1 real-time data, and D2 is compared to Channel 2 real-time
data. Comparison data must either be measured, then stored into data registers D1
and D2, or recalled into registers D1 and D2 from disk or non-volatile RAM. For
instructions on how to store spectra to disk, see the section “Storing Data to a Disk
File.” For instructions on how to store spectra to data registers, see the section
“Storing Data in Data Registers.”
For this section, the comparison data is assumed to be stored in the disk file
COMPARE1.DAT” and “COMPARE2.DAT.” To recall these spectra into D1 and
D2, respectively, follow these steps:
Insert the 3.5” disk containing the files “COMPARE1.DAT” and
COMPARE2.DAT.”
Press the [Save/Recall] hardkey.
Press the [RECALL DATA] softkey (F4).
Press the [RCL TRACE AND SCALE] softkey (F1).
Press the [FROM FILE INTO D1] softkey (F1).
Press the [CLEAR ENTRY] softkey (F9).
Enter COMPARE1.DAT and press the [ENTER] softkey (F1).
Data from the file “COMPARE1.DAT” now appears in data register D1. Next, recall
COMPARE2.DAT” into dataregister D2 as follows:
Press the [Save/Recall] hardkey.
Press the [RECALL DATA] softkey (F4).
Press the [RCL TRACE AND SCALE] softkey (F1).
Press the [FROM FILE INTO D2] softkey (F2).
Press the [CLEAR ENTRY] softkey (F9).
Enter COMPARE2.DAT and press the [ENTER] softkey (F1).
Recalling the Spectra from Non-Volatile RAM
The comparison data files “COMPARE1.DAT” and “COMPARE2.DAT” are stored in
the non-volatile RAM. To recall these spectra into D1 and D2, respectively, follow
these steps:
Press the [Save/Recall] hardkey.
Press the [DEFAULT DISK] softkey (F9).
Press the [NON VOL RAM DISK] softkey (F1).