Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
6
Any frequency below 51.2 kHz may of course be entered, however the analyzer will
select the lowest span that contains the desired frequency. For example, if 1200 Hz
is entered, the analyzer is switched to a span of 1600 Hz.
* Available in single-channel mode when anti-alias filter is turned off.
To begin data acquisition,
Press the [Start] hardkey.
To automatically scale the display to fit the data,
Press the [Scale] hardkey.
Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Frequency Spans Available
for the AGILENT 35670A
102.4 kHz * 100 Hz
51.2 kHz 50 Hz
25.6 kHz 25 Hz
12.8 kHz 12.5 Hz
6.4 kHz 6.25 Hz
3.2 kHz 3.125 Hz
1.6 kHz 1.5625 Hz
800 Hz 0.78125 Hz
400 Hz 0.390625
200 Hz 0.1953125 Hz