Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
46
Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted..
Press the [Active Trace] hardkey.
Press the [B] softkey (F1)
Press the [Meas Data] hardkey.
Press the [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
Press the [TIME CHANNEL 2] softkey (F5)
Press the [Scale] hardkey.
Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Set up triggering on Channel 1 to 5% of range:
Press the [Trigger] hardkey.
Press the [CHANNEL 1 234] softkey (F3) until 1 is highlighted.
Press the [TRIGGER SETUP] softkey (F6)
Press the [CHANNEL LEVEL] softkey (F1)
Enter 5 then press the [PERCENT (%)] softkey (F4).
Press the [ALL CHANNELS] softkey (F8).
Press the [CHANNEL * DELAY] softkey (F9).
Enter -0.01 then press the [S] softkey (F1).
Mount the accelerometers on top of each other or very close together on a light test
structure, then:
Press the yellow [Start] hardkey.
Tap the structure in the area around the two accelerometers to trigger a
measurement.
If the measurement overloaded, then tap again after the range has
increased. Tap again until a trace is not overloaded.
Press the [Marker] hardkey.
Toggle the [COUPLED ON OFF] softkey (F2) until ON is highlighted.
Examine the two time traces with the coupled markers to determine if the
accelerometers are in phase or out of phase.
Note whether the accelerometers are in phase or out of phase. The function to
determine differential amplitude depends on this relationship.
Measuring Amplitudes, Differential Amplitude, and Phase
For machinery motion measurements, one display setup can show:
displacement of Channel 1 in mils peak-to-peak
displacement of Channel 2 in mils peak-to-peak