Then set up trigger parameters (for a positive event, positive slope) as follows:
Press the [TRIGGER SETUP] softkey (F6)
Press the [CHANNEL LEVEL] softkey (F1)
Enter 5 then press the [PERCENT (%)] softkey (F4)
Press the [SLOPE POS NEG] softkey (F5) until POS is highlighted.
Trigger Delay
Both channels should acquire data using the same pre-trigger delay. To set up a
0.01s pre-trigger condition, follow these steps:
Press the [Trigger] hardkey.
Press the [TRIGGER SETUP] softkey (F6)
Press the [ALL CHANNELS] softkey (F8)
Press the [CHANNEL * DELAY] softkey (F9)
Enter -.01 then press the [S] softkey (F1)
To start the data acquisition,
Press the yellow [Start] hardkey.
The analyzer should show in highlighted text above the trace:
WAITING FOR CH 1 TRIGGER