Section 6 : Measuring a Single Channel Time Waveform
This section contains procedures for setting up and measuring a single channel time
waveform using an AGILENT 35670A Dynamic Signal Analyzer. Included are
steps to select single channel operation, set up the trace time length, and set up
triggering to capture a transient event. Instructions will also be presented to scale the
time trace.
Selecting Single Channel Operation
To measure a single-channel time waveform, turn on the AGILENT 35670A and
then:
Press the [Inst Mode] hardkey
Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
Note: if the AGILENT 35670A was already turned on and in an unknown
state, then preset the analyzer by doing the following:
Press the green [Preset] hardkey
Press [DO PRESET] softkey (F1).
Then set the instrument mode to single channel as above.
Time Record Length in Seconds vs. Resolution
Span - Hz 100 lines 200 lines 400 lines 800 lines 1600 lines
51200 Hz
0.001953 s 0.003906 s 0.007813 s 0.015625 s 0.03125 s
25600 Hz
0.003906 s 0.007813 s 0.015625 s 0.03125 s 0.0625 s
12800 Hz
0.007813 s 0.015625 s 0.03125 s 0.0625 s 0.125 s
6400 Hz
0.015625 s 0.03125 s 0.0625 s 0.125 s 0.25 s
3200 Hz
0.03125 s 0.0625 s 0.125 s 0.25 s 0.5 s
1600 Hz
0.0625 s 0.125 s 0.25 s 0.5 s 1 s
800 Hz
0.125 s 0.25 s 0.5 s 1 s 2 s
400 Hz
0.25 s 0.5 s 1 s 2 s 4 s
200 Hz
0.5 s1 s2 s4 s8 s
100 Hz
1 s2 s4 s8 s16 s
50 Hz
2 s 4 s 8 s 16 s 32 s
25 Hz
4 s 8 s 16 s 32 s 64 s
12.5 Hz
8 s 16 s 32 s 64 s 128 s
6.25 Hz
16 s 32 s 64 s 128 s 256 s
3.125 Hz
32 s 64 s 128 s 256 s 512 s
1.5625 Hz
64 s 128 s 256 s 512 s 1024 s
0.78125 Hz
128 s 256 s 512 s 1024 s 2048 s
0.390625 Hz
256 s 512 s 1024 s 2048 s 4096 s
0.1953125 Hz
512 s 1024 s 2048 s 4096 s 8192 s