Agilent Technologies Agilent 35670A Stereo System User Manual


 
AGILENT 35670A
Supplemental Operator’s Guide
51
Attach one end of stinger to shaker and one end to force transducer
Connect amplifier output to shaker
Connect amplifier input to AGILENT 35670A source
Connect the transducers
Connect force transducer to Channel 1
Connect accelerometer to Channel 2
Preset the AGILENT 35670A
Turn on the AGILENT 35670A, or
Press the green [Preset] hardkey
Press the yellow [Pause/Cont] hardkey if desired.
Enter Channel 1 input parameters
Press the [Input] hardkey
Press [CH1 AUTO UP ONLY] softkey (F5).
Press [FRONT END CH1 SETUP] softkey (F7).
Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
Press the [Input] hardkey
Press [XDCR UNIT CH1 SETUP] softkey (F8)
Press the [XDCR SENSITVTY] softkey (F6).
Enter 8, then press the [mV/EU] softkey (F2).
Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
Press the [XDCR UNIT LABEL] softkey (F7).
Press the [MORE CHOICES] softkey (F9).
Select the [lb] softkey (F6).
Enter Channel 2 input parameters
Press the [Input] hardkey
Press [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
Press [CH2 AUTO UP ONLY] softkey (F5).
Press [FRONT END CH2 SETUP] softkey (F7).
Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
Press the [Input] hardkey
Press [XDCR UNIT CH2 SETUP] softkey (F8)
Press the [XDCR SENSITVTY] softkey (F6).
Enter 10.2, then press the [mV/EU] softkey (F2).
Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
Press the [XDCR UNIT LABEL] softkey (F7).
Select the [g] softkey (F2).