VXI VT1433B Stereo Receiver User Manual


 
Parameter Lists
This section shows which parameters are global parameters, which are
channel-specific and what types of channels the channel-specific parameters apply
to. Default values are shown for all of these parameters. In addition, each
parameter is categorized as “abort”, “wait”, “immediate” or “glitch” depending on
the behavior when this parameter is changed during a running measurement.
Those with “abort” cause the measurement to abort. Those with “wait” don’t take
effect until the start of the next measurement. Those with “immediate” take
effect immediately. Those with “glitch” take effect immediately and may cause
glitches in the data that is read back or on the source output if the parameter is
applied to a source channel.
Global Parameters
Parameter Default Value Changes
append_status Off Immediate
arm_channel None Immediate
arm_mode Auto Arm Immediate
arm_time_interval 1 s Immediate
auto_group_meas On Wait
avg_mode None Wait
avg_number 10 Wait
auto_trigger Auto Trigger Abort
avg_update 10 Wait
avg_weight 1 Immediate
blocksize 1024 Abort
cal_dac 0 Immediate
cal_voltage 0 V Immediate
calin Grounded Immediate
center_freq 2 kHz Immediate
clock_freq 51.2 kHz Abort
clock_master Off Abort
clock_source Internal Abort
data_mode Block Mode Abort
data_port VME Abort
data_size 16 Bit Integer Abort
VT1433B User's Guide
The Host Interface Library
4-5