Performance Verification
Performance Verification
4Ć10
CSA 803A Test Record (Cont.)
MaximumOutgoingIncomingMinimumPerformance Test
Vertical Offset Accuracy, Fourth Compartment "2mV
Step 10. 400 mV 398 mV __________ __________ 402 mV
Step 12. 1 V 0.998 V __________ __________ 1.002 V
2 V 1.998 V __________ __________ 2.002 V
-400 mV -402 mV __________ __________ -398 mV
-1 V -1.002 V __________ __________ -0.998 V
-2 V -2.002 V __________ __________ -1.998 V
Sweep Rate Accuracy
ns Sweep Rate Accuracy
Step 10. 2 ns/div ("0.05%) 18.172 ns __________ __________ 18.192 ns
Step 14. 1 ns/div ("0.08%) 9.083 ns __________ __________ 9.098 ns
ps Sweep Rate Accuracy
Step 18. 100 ps/div ("0.8%) 902 ps __________ __________ 916 ps
Step 28. 10 ps/div ("2.5%) 248.1 ps __________ __________ 260.5 ps
Step 40. 1 ps/div ("10%) 23.1 ps __________ __________ 28.3 ps
Internal Clock
Step 7. Rise Time __________ __________ 2.5 ns
Step 9. Frequency 97 kHz __________ __________ 103 kHz
Duty Cycle 47% __________ __________ 53%
Calibrator Output
Step 7. Rise Time __________ __________ 250 ps
Amplitude 225 mV __________ __________ 275 mV