Specifications
CSA 803A Service Manual
1Ć3
TableĂ1Ć1:ăMeasurement Limits, Specifications, Adjustments, and Functional Tests (Cont.)
Part and Description Functional
Tests
Adjustments
(Adjust)
Specifications
(Check)
Measurement
Limits (Examine)
Sweep Rate Accuracy 0.05% at 2 ns/div
0.08% at 1 ns/div
0.8% at 100 ps/div
2.5% at 10 ps/div
10% at 1 ps/div
none none yes
Triggering, External Direct yes
300 MHz Sensitivity 100 mV
pĆp
at 300 MHz 100 mV stable
display
none
1000 MHz Sensitivity 100 mV
pĆp
at 1000 MHz 100 mV stable
display
none
3.0 GHz Sensitivity 100 mV
pĆp
at 3.0 GHz 100 mV stable
display
none
Prescaler Triggering 600 mV
pĆp
at 2 GHz 600 mV stable
display
none yes
ą ą 600 mV
pĆp
at 8 GHz 600 mV stable
display
none
ąą 2V
pĆp
at 10 GHz 2 V stable display none
Internal Clock
yes
Rise Time ≤2.5 ns ≤2.5 ns none
Frequency 100 kHz ±3% 100 kHz ±3% none
Duty Cycle 50% ±3% 50% ±3% none
Calibrator Output yes
Rise Time ≤250 ps (20% - 80%) ≤250 ps
(20% - 80%)
none
PeakĆtoĆpeak Amplitude 250 mV ±10% 250 mV ±10% none