Tektronix CSA 803A Stereo System User Manual


 
Adjustments
CSA 803A Service Manual
5Ć3
Table 5Ć2 (Test Equipment), contains suggested test equipment for use in
this section. Procedure steps are based on the test equipment examples
given, but you may substitute other equipment with similar specifications.
The use of different equipment may alter test results, setup information, and
related connectors and adapters.
TableĂ5Ć2:ăTest Equipment
Description
Minimum
Specification
Examples of Applicable
Test Equipment
Digital Voltmeter
(w/test leads)
v0.005% Accuracy FLUKE 8842A Digital Voltmeter
Frequency Counter One Part in
1,000,000 Accuracy
Tektronix DC 5010 Universal
Counter/Timer with a
TM 5000ĆSeries Power Module
Test Terminal any GPIB
(IEEEĆ1978) controlĆ
ler, or ASCII termiĆ
nal equipped with
an RSĆ232ĆC port.
Requires compatĆ
ible RSĆ232ĆC serial
interface cable
IBM compatible PC with
terminal emulation software
Coaxial Cable, 50 W
(3 required)
50 W, 36Ćinch, male
BNC connectors
Tektronix Part 012-0482-00
Coaxial Cable, RF RF cable, 12 in.,
SMA connectors
Tektronix Part 174-1364-00
RF cable, 8.5 in.,
SMA connectors
Tektronix Part 174-1120-00
Adapters BNC femaleĆtoĆmale
Dual Banana
Tektronix Part 103-0090-00
BNC maleĆtoĆfemale
Dual Banana
Tektronix Part 103-0035-00
SMA maleĆtoĆBNC
female (8 required)
Tektronix Part 015-1018-00
SMA femaleĆtoĆBNC
male
Tektronix Part 013-0126-00
SMA maleĆtoĆN
male
Tektronix Part 015-0369-00
Attenuator, 10X 20 dB attenuation,
50 W, one male and
one female
Tektronix Part 011-0059-02
Test Equipment