Performance Verification
Performance Verification
4Ć8
CSA 803A Test Record (Cont.)
MaximumOutgoingIncomingMinimumPerformance Test
Vertical Gain, Second Compartment "1.2% at Checkpoint
Step 11. 5 mV/div -20.5 mV __________ __________ -19.5 mV
10 mV/div -41.0 mV __________ __________ -39.0 mV
20 mV/div -82 mV __________ __________ -78 mV
50 mV/div -205 mV __________ __________ -195 mV
100 mV/div -410 mV __________ __________ -390 mV
200 mV/div -820 mV __________ __________ -780 mV
5 mV/div 19.5 mV __________ __________ 20.5 mV
10 mV/div 39.0 mV __________ __________ 41.0 mV
20 mV/div 78 mV __________ __________ 82 mV
50 mV/div 195 mV __________ __________ 205 mV
100 mV/div 390 mV __________ __________ 410 mV
200 mV/div 780 mV __________ __________ 820 mV
Vertical Offset Accuracy, Second Compartment
"2mV
Step 10. 400 mV 398 mV __________ __________ 402 mV
Step 12. 1 V 0.998 V __________ __________ 1.002 V
2 V 1.998 V __________ __________ 2.002 V
-400 mV -402 mV __________ __________ -398 mV
-1 V -1.002 V __________ __________ -0.998 V
-2 V -2.002 V __________ __________ -1.998 V
Vertical Gain, Third Compartment "1.2% at Checkpoint
Step 11. 5 mV/div -20.5 mV __________ __________ -19.5 mV
10 mV/div -41.0 mV __________ __________ -39.0 mV
20 mV/div -82 mV __________ __________ -78 mV
50 mV/div -205 mV __________ __________ -195 mV
100 mV/div -410 mV __________ __________ -390 mV
200 mV/div -820 mV __________ __________ -780 mV