Emerson 1700 Satellite Radio User Manual


 
62 Micro Motion
®
Model 1700 and 2700 Transmitters
Specifications
A.1.4 Power supply
The Model 1700/2700 transmitter’s power supply:
Has a self-switching AC/DC input
Complies with low voltage directive 2006/95/EC per EN 61010-1 (IEC 61010-1) with
amendment 2
Meets Installation (Overvoltage) Category II, Pollution Degree 2 requirements
Has an IEC 127–1.25 slowblow fuse
AC power requirements
85 to 265 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum
DC power requirements
18 to 100 VDC, 6 watts typical, 11 watts maximum
At startup, transmitter power source must provide a minimum of 1.5 amps of short-term
current at a minimum of 18 volts at the transmitter power input terminals
Minimum 22 VDC with 1000 feet (300 meters) of 18 AWG (0,8 mm
2
) power supply cable
A.1.5 Environmental requirements
Ambient temperature limits
Operation: –40 to +140 °F (–40 to +60 °C)
Storage: –40 to +140 °F (–40 to +60 °C)
Display responsiveness decreases, and display may become difficult to read, below –4 °F (–20 °C).
Above 131 °F (55 °C), some darkening of display may occur.
If possible, install the transmitter in a location that will prevent direct exposure to sunlight.
Different ambient temperature requirements may apply when installing the transmitter in a hazardous
area. Refer to the approval documentation shipped with the transmitter or available on the Micro
Motion web site.
Humidity limits
Relative humidity between 5 and 95%, non-condensing at 140 °F (60 °C)
Vibration limits
The transmitter meets IEC 68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1.0 g
A.1.6 Ambient temperature effect
On analog outputs ±0.005% of span per °C
A.1.7 EMC compliance
The transmitter complies with the following EMI effects standards:
NAMUR NE21 (May 1999) with the exception of Voltage Dip when powered by 24 VDC
Complies with EMC directive 2004/108/EC per EN 61326 Industrial